MANIFill Flow
MANIFill Flow
The light-curing MANIFill Flow composites – easy handling thanks to the different types of viscosity. MANIFill Flow 55% filler content with low viscosity for direct, minimally invasive restorations.

Product Characteristics
- Easy and comfortable application
- No dripping – easy and highly accurate application
- Excellent physical properties
- Naturally aesthetic restorations
- Thixotropic behavior for precise placement
Indications
Light-curing, flowable composite with low viscosity for direct, minimally invasive restorations.
- Fissure sealing
- Class V fillings
- Minimally invasive Class I, II, and III fillings
- Correction of enamel defects
- Blocking out undercuts
- Minor shape and shade corrections on enamel

Step by Step
Preparation of the tooth surface
- Clean the tooth structure with a fluoride-free polishing paste.
Cavity preparation
- Preparation of the cavity according to the general rules of the adhesive technique.
Protection of deep cavities
- In the case of very deep cavities close to the pulp, cover the corresponding areas with a calcium hydroxide preparation
Matrix application for approximal defects
- For cavities with approximal parts, apply and fix a transparent matrix.
Etching and bonding
- Etch and bond according to the manufacturer’s instructions
Application of the flow composite
- Apply Flow in thin layers (max. 2 mm) directly into the cavity, do not introduce any air bubbles when applying with the cannula.
- Cure each applied layer with a commercially available polymerization device for 40 seconds. During polymerization, a dispersion layer forms on the surface, which must not be touched or removed. It is used to connect the subsequent composite layers.
- ManiFill Flow can be finished and polished immediately after polymerization. Finishing diamonds, flexible discs, silicone polishers and polishing brushes are suitable for finishing. A light-curing unit with an emission spectrum of 350 – 500 nm should be used for polymerization.
Light intensity for curing | ≥ 650 mW / cm2 |
Wavelength for curing | 350 – 500 nm |
Curing time | 40 sec. |
Maximum layer thickness | 2 mm |
- Store at 10 – 25 °C (50 – 77 °F). The material should have reached room temperature before use. Close the syringe tightly immediately after use.
- The maximum shelf life is printed on the label of the respective syringe. Do not use after the expiry date!
- Undesirable side effects of this medical device are extremely rare if processed and used properly.
Vickers Hardness | 374 MPa |
---|---|
Bending strength | 134 MPa |
E-module | 6998 MPa |
Compressive strength | 403 MPa |